Where should I put the electrodes for EMS?

Electrode placement for EMS (Electrical Muscle Stimulation) is very important for obtaining the best results. Muscle activation works best when the motor nerve is located and stimulated.

Usually the Bipolar Technique is used, in which two electrodes are placed over the bulk of the muscle - with one electrode (usually the black pin) placed over the muscle’s motor point (also used in Asymmetrical Biphasic Pulse Stimulation).

The motor point is the area on the skin that is located closest to the muscle’s motor nerves entry point. Here, it is easiest to trigger a contraction by electrical stimulation.

Localising the motor point provides the cleanest and most comfortable contractions. These points can be located by using a diagram or muscle map such as the one below (see bottom of the page).

Move the electrode across the skin and locate the point over the muscle that gives the cleanest contraction. Large muscle groups may require stimulation with two channels i.e. using four electrodes simultaneously.

Electrode Pad Care

The pads which are supplied with TensCare E.M.S. units are self-adhesive and can be used several times. Skin must be allowed to breathe - therefore the pads should be removed periodically. When not in use, the pads should be placed on the pad holder or returned to the clear plastic shield.

The condition of the pads does affect the conductivity and performance of the unit. When the pads initially lose their adhesive quality, it is possible to reactivate their adhesiveness by applying a fine spray of water. Once the pads have finally lost their adhesive quality, replacement electrodes should be purchased.